Dual O2/CO2 Monitor, 0-25% and 0-10,000ppm
Description
Dual O2/CO2 Monitor, 0-25% and 0-10,000ppm
PureAire Monitoring Systems’ Dual O2/CO2 Monitor is designed for continuous monitoring of oxygen and carbon dioxide in confined spaces. The monitor is ideal for facilities that use inert gases, including, but not limited to, nitrogen, helium, carbon dioxide, and argon.
The Dual O2/CO2 Monitor is suitable for environments such as cryogenic facilities, breweries, food processing plants, pharmaceutical manufacturing, laboratories, hospitals, and universities. The monitor will remain accurate at temperatures as low as -40c. PureAire’s durable, non-depleting, zirconium oxide sensor cell and non-dispersive, infrared sensor (NDIR) cell, will last 10+ years in a normal environment, without needing to be replaced.
Sampling Method & Range | Diffusion, 0-25% O2, 0-10,000ppm CO2 |
Accuracy | Delivers ±0.2% O2 accuracy (±1% of full range- O2) ±30 ppm, ± 3% of measured value – CO2 |
Operating Temperature | -0 to + 60° C |
Display | ¾” backlit LCD digital display |
Sensor Type | Long-life zirconium oxide sensor for O2 and NDIR for CO2 |
Alarm indicators | Built-in Audible 90dB horn, Orange LED alarm indicator for CO2, Red LED alarm indicator for O2 |
Signal Outputs | One user-adjustable Alarm relay for Oxygen and one user-adjustable Alarm relay for CO2. Relays are rated 2 amp 24VDC / 24VAC |
Analog Output | Analog 4-20mA output (Active), selectable for either Oxygen or CO2. Optional 0-5 VDC output for CO2. |
Power Requirements | 24VDC 350mA |
Dimensions | 6.5 (W) x 4.5 (H) x 3.25 (D) inches; (165.1 mm x 114.3 mm x 82.5 mm) |
Weight | 1.6 lbs. (.8 kg) |
Enclosure | Polycarbonate |
Required calibration | Calibration is not required for Oxygen. None for CO2 when Auto Cal function is activated. Periodic span adjustment as needed for Oxygen. |
Country of Origin | USA |
Harmonized Tariff Code | 8531.10.0045 |
PureAire